CENTRIFUGE 800 Series
CENTRIFUGE
800 Series
Instruction Manual
Catalogue
II. Principle
III. Safety Precautions
IV. Operation Instruction
V. Maintaining
I. Introduction
Desktop centrifuge is a routine laboratory equipment, widely used in biological, chemical, pharmaceutical and other research and education and production sector. Low-speed desktop centrifugal maximum speed of 4000r / min, suitable for radioimmunoassay and separation of cells and large particles, there are two kinds of regular and irregular.
II. Principle
Model |
Rotation Speed |
Volume |
Maximum Relative Centrifugal force |
Power supply |
Time Range |
800 |
0-4000rpm |
20ml×6 |
1430xg |
200V,50HZ |
|
800D |
0-4000rpm |
20ml×6 |
1430xg |
200V,50HZ |
0~60min |
80-3 |
0-4000rpm |
20ml×6 |
1430xg |
200V,50HZ |
0~60min |
80-1 |
0-4000rpm |
20ml×6 |
1430xg |
200V,50HZ |
0~60min |
80-2B |
0-4000rpm |
20ml×12 |
2325xg |
200V,50HZ |
0~60min |
80-2 |
0-4000rpm |
20ml×12 |
2325xg |
200V,50HZ |
0~60min |
90-1 |
0-4000rpm |
20ml×6 |
1430xg |
200V,50HZ |
0~60min |
90-2 |
0-4000rpm |
20ml×12 |
2325xg |
200V,50HZ |
0~60min |
III. Safety Precautions
1. It is forbidden to run without load.
2. Before running, check the rotor for corrosion damage.
3. Do not open the centrifuge cover while the rotor is running, do not touch the running rotor.
4. It is forbidden to run in an unbalanced load condition.
IV. Operation Instruction
1. The centrifuge table should be flat and solid, the work should be neatly clean, dry and well ventilated, the ambient temperature to 5-32 ° C is appropriate.
2. Open the centrifuge cover, wipe the cavity and turn the head clean.
3. Connect the external power supply.
4. Assemble the tube with the same consistency.
5. Adjust the speed control knob to the desired speed. (800-1 type need to hit the timer normally open ON position, or hit the timing range 0-60 minutes)
6. Before each stop, the speed knob must be placed in the minimum position "O". Cut off the power.
7. Wipe the inner cavity and turn the head and close the centrifuge cover.
V. Maintaining
1. Do not place anything on the centrifuge cover. After each use, be sure to clean the cavity and turn the head, cover the centrifugal cover to prevent dust, dirt into.
2. If the centrifuge is not used for a long time, before the re-use, the centrifuge cover should be opened for some time, dry the cavity.
3. The carbon brush of the motor after the use of 3000 hours, should be replaced in time to avoid wear commutator.
4. Centrifuge by the long-term use, natural wear is a normal phenomenon. After a year of use, if the failure or difficulties, the enterprises will continue to provide quality services, to help deal with.