Clean type Pass Box USER’S MANUAL
Clean type Pass Box USER’S MANUAL
I. Application
Pass box is a kind of air purification equipment that used in the clean room.Clean type pass box is used for delivery small items between clean room and non clean areas.Using the pass box can effectively decrease the number of open the door for clean room,and reduce the pollution of the clean room to a minimum.
ⅡStructure Characters
1. The working area is made of 304 stainless steel,the box body is made of 304 stainless steel,smooth and clean.
2. On both side of the door with electrical interlock device to realize interlock.When we use the machine,both sides of the door can not be open at the same time.
Electrical interlock device:Internal use of integrated circuit,electromagnetic locks,control panel,lamp and so on to achieve chain.When one door is opened,the switch indicator light of the other one is not bright,to inform that the door can not be open.At the same
1. me,the electromagnetic action implementation chain lock.When the door is closed,the electromagnetic lock of the other door is starting to work,indicator lights up at the same time,the other door will open.There is a hole at the bottom of the control panel,the hole in the condition of the no electricity or emergency,with 1-3 mm wire sticking to the bottom,gently moves in the direction of the door can be directly open the door.
2. The complete set of the electric control is in the internal side of the pass box.If there are obstacles of the circuit can remove the inside of the plate for inspection or servicing.
3. Pass box is equipped with special sealing strip,to ensure the tightness according to different use increase indoor lights and germicidal lamp.
III Manual for the controller
The system operation panel set two groups the A0 and A1 (a) each door, each group separately on the operation panel with seven buttons (power button, the door key, set the key, sterilization, fan keys, raised key, cut key).Set up before use sterilization, fan, lighting time (0-999 seconds or 0-999).The original set to sterilization for 20 seconds, air supply for 30 seconds, lighting 0 seconds.
Time setting: (such as the need to set up 40 seconds, sterilization supply 50 seconds, lighting 10 seconds) hold the set key for 3 seconds and then let go, set a key indicator, sterilization indicator lights flashing at the same time, press adjust up and down keys to adjust time to 40;Press the set key again (at this point does not need to hold 3 seconds), air supply indicator lights flashing, press adjust up and down keys to adjust time to 50;Then press set up (at this point does not need to hold 3 seconds), the power indicator light flashing (set) for lighting time at this time, press adjust up and down keys to adjust time to 10.Note: the setup time must be in the case of setting indicator set three groups of time to complete, otherwise is invalid.
In standby state, according to "product" take turns beating, the power indicator, when the power was turned off, the power indicator light flashing.
Deliver goods when press the door key to open the door (open the door for more than 5 seconds, buzzer alarm) at this time the door indicator, the other side to open the door indicator lights flashing (at this time can't open the door), put the goods are door closed all the lights go on both sides, sterilization lamp or fan to start working at this time.Work LED display supply time 50 began to countdown.After the ending of sterilization time and display lighting Air shower time.Three groups of time preference according to the order for the fan, sterilization, lighting.Fan, sterilizing panel indicator synchronous on at work.
Light switch is controlled by the door, when any lighting the lamp when the door opened, closed after start the countdown.
Work mode setting: in the standby time press set key to open the door indicator (long by 10 seconds), let go of the set key at this time, set the indicator, enter setup mode, according to the SEC (in seconds), press raised shows FEN said (in minutes) for the unit, has chosen the SEC for 0-999 seconds, such as the need to choose specific provisions of 0-999 FEN;Press the set key again shows "1 -" pattern 1, press adjust up and down keys to choose working mode;Then press the set key, sterilization indicator lights flashing, press adjust up and down, sets the sterilization time (if 30);Light is flashing, press the set key, fan set supply after the adjust up and down button (if 20);Then press the set key, the power indicator light flashing (set) for lighting time at this time, press adjust up and down keys set lighting time (if 10).Note: when set all setup process must be completed, otherwise is invalid.Note: the factory Settings for the pattern 8
Working mode table:
manual
manual
One-way automatic
One-way automatic
One-way automatic
Two-way automatic
Two-way automatic
Two-way automatic
Panel time synchronization
★
★
★
★
★
Panel time are not synchronized
★
★
★
Open the fan work is invalid
★
★
Open the sterilization work is invalid
★
★
Fan sterilization work to open the door is invalid
★
★
★
VI Technical Parameters
Pattern 1
Model 2
Pattern 3
Pattern 4
Model 5
Model 6
Mode 7
Model 8
VII Common faults and maintenance methods
Fault Phenomenon
Fault Cause
Methods
The door light is not bright
The opposite interlock is bad
Change the interlock
When the door light is Off,the door not open
This side interlock is bad
Change the interlock
The door light still on,the door not open
This side interlock is bad
Change the red buttom
VIII .Electrical chart