Glass bead sterilizer
Y09-3016 particle counter add Bluetooth printing function
Overview:
The product uses a color LCD touch screen and is the first hand-held dust particle counter in China.
Technical Parameters:
1. Meet the JJF-1190-2008 dust particle counter verification specifications:
2, self-cleaning time ≤ 10min
3. Measurement repeat relative deviation ≤ ± 10%
4, particle size distribution error 0.5um 5um ≤ ± 20%
5, particle concentration indication error 0.5um ≤ ± 20%
6, sampling flow and error: 2.83L / min ≤ ± 5%
7, measurable particle size range: 0.3, 0.5, 1, 3, 5, 10um, divided into 6 particle size channels, 0.3um signal to noise ratio ≥ 2.5: 1
8, the maximum measurable particle concentration: ≤ 30K / L
9, current DC7.2V, 1800mAh
Performance parameter
Test particle size |
0.3, 0.5,1,3,5,10μm |
Light source |
Laser diode (lifetime greater than 30,000 hours) |
Flow |
2.83 L/M(0.1cfm) |
Counting efficiency |
50±10%@0.3μm, 100±10%@0.5μm |
Measuring repeated relative deviation |
≤±10% |
sampling time |
User-defined (1 minute to 99 minutes)) |
Sampling delay |
User-defined (1 minute to 99 minutes) |
Sampling frequency |
1 to 99 times or continuously |
Alarm level |
Class100 to Class 300,000 (FED209E standard) or 5 to 9 (ISO14464-1 standard) |
test method |
Single / repeat / continuous / timing / average |
Data memory capacity |
1000 sets of measurement data |
Error indication |
xceeding the maximum count, the laser power is attenuated, exceeds the calibration flow (>±5%), and the battery is low. |
Communication Interface |
USB |
Maximum communication rate |
12MBPS |
Charging time |
3 hours, 12 hours for the first charge |
Self-cleaning time |
≤10min |
Maximum concentration |
"+/-5% at 2000000 particles/ft3 (71000 Particle/L) |
Screen display mode |
LCD touch screen |
printing method |
External printer with Epson ribbon |
Sampling position number |
1000 groups |
Counting mode |
Cumulative number, interval, |
power supply |
Rechargeable Ni-MH battery (6V/2500mAH) or AC adapter (input 100~240V) |
Battery working time |
Continuous test time is greater than 6 hours |
Dimensions |
220*40*120mm |
Weight |
1kg |
Environmental conditions |
Working environment 15~35°C, 10-85%RH; storage environment: -10~50°C, 10-85%RH |
Standard accessories |
AC adapter, battery charger, clear filter, power sampling head, portable protection box, operating instructions, original calibration report, Bluetooth printer, data export software |
Optional accessories |
|
Verification standard |
JIS-B-9921(1997), ASTM-F649-01, ASTM-F328-98 (NIST可溯源) JJG547-1988 GB/T6167-2007 JJF1190-2008 |
UCL criteria |
ISO14644 FS209E |