ARTICLE DETAILS

Y09-301AC Y09-301(AC/DC) Particle Counter

日期:2024-10-08 17:50
浏览次数:301
摘要: Y09-310AC-DC (Battery Operated)28.3L/minLaser Particle Counter is used to detect the particles in the cleanroom. It can measure the class 100 to 300 k environment. The instrument uses the laser diode as the light source, LCD-panel displays, six different size ranges simul taneously. The built-in printer can print count results. The counter can store 1000 samples data. Software creates data files for viewing or export to EXCEL sheet to computer. The instrument is widely used in t

Y09-310AC-DC (Battery Operated)28.3L/minLaser Particle Counter

is used to detect the particles in the cleanroom. It can measure the class 100 to 300 k environment. The instrument uses the laser diode as the light source, LCD-panel displays, six different size ranges simul taneously. The built-in printer can print count results. The counter can store 1000 samples data. Software creates data files for viewing or export to EXCEL sheet to computer. The instrument is widely used in the fields of micro-electronics, precision machinery, optics, pharmacy and sanitation, biology, food and beverage, aerospace and so on.

Specifications

Weight

7.5Kg

Max. Consumption

48W

Power Source

DC 16.8V (Lithium Ion Battery)

Size Channels

0.3, 0.5, 1, 3, 5, 10 um

Calculation

95%UCL calculation, Display Particle Concentration

Flow Rate

28.3L/min (Imported Pump)

Sample Time

1---10min (Selectable)

Zero Count

10min

Location Labels

0-999

Printer

Built-in Thermal Printer

Alarm

10,100, 1000, 10000 ,100000,300000Class /ISO

Data Storage

1000 Samples (With PC Software)

Light Source

Laser Diode (30, 000 hours Mean Time Between Failures)

Maximum Rated Particle Concentration

35000 particles/L

Operating Time

3 hours

Environmental Temperature

10degrees celsius-35degrees celsius

Humidity

20%-75%RH

Atmosphere Pressure