摘要: BHC-1300IIA/B2 Clean Biological Safety Cabinet Clean biological safety cabinet is necessary in animalcule lab, especially in the condition needing protection measure, such as medicine, pharmacy, scientific research and so on. Main characters 1 Air insulation design to avoid cross pollution inside and outside air circulation within 30% ,70% discharge, vertical laminar negative pressure. 2 Equipped with spring up/down movable door in the front and back of work bench

BHC-1300IIA/B2 Clean Biological Safety Cabinet

      Clean biological safety cabinet is necessary in animalcule lab, especially in the condition needing protection measure, such as medicine, pharmacy, scientific research and so on.
Main characters

1 Air insulation design to avoid cross pollution inside and outside air circulation within 30% ,70% discharge, vertical laminar negative pressure.

2 Equipped with spring up/down movable door in the front and back of work bench, flexible and convenient to locate

3 Equipped with special filter on ventilation to keep vented air conforming to national standard.

4 Contact switch adjusts voltage to keep wind speed in working area in ideal state all the time.

5 Operate with LED panel.

6 The material of the work area is 304 stainless steel.

No. Index/Model

BHC-1300IIA/B2 Clean Biological Safety Cabinet

Cleaning level

Grade 100(209E U.S. Federal)

Number of bacteria

≤0.5per utensil.hour(90mm utensil)

The average wind speed- inner side

0.38+-0.025m/s

The average wind speed- middle

0.26+-0.025m/S

The average wind speed-inside

0.35+-0.025m/s

Air tightness

<=10-6m/S 500Pa leakage rate

Noise

≤62dB

Power supply

AC, 220 V/50Hz

Illumination

≥300LX

Half ventilation peak value

≤ 5μm(XYZ direction)

Max. power consumption

600W

Weight

250KG

Dimension of working area(W1×D1×H1)

1000*700*680mm

Overall dimension W*D*H

1200*780*1950mm

Efficient filter specifications and quantities

995*640*38*1      

Specification and number of light from
firefly/ ultraviolet light

16W*2/15W*2

Suited number

 Double single

 Note: Special specification of non-standard type clean bench is also available on demand.