NEWS DETAILS

Installation process description of laminar flow cabinet

日期:2025-05-29 13:52
浏览次数:255
摘要:Ultra clean workbench is a purification equipment that provides a locally dust-free and sterile working environment, widely used in biological laboratories, pharmaceutical production, scientific research and other fields. The correct installation process is crucial for ensuring the normal operation and service life of the ultra clean workbench. The following is a description of the installation process for the ultra clean workbench: 1. Preparation stage -Before installation, it is necessary
Ultra clean workbench is a purification equipment that provides a locally dust-free and sterile working environment, widely used in biological laboratories, pharmaceutical production, scientific research and other fields. The correct installation process is crucial for ensuring the normal operation and service life of the ultra clean workbench. The following is a description of the installation process for the ultra clean workbench:
1. Preparation stage
-Before installation, it is necessary to ensure that the outer packaging of the workbench is intact and undamaged. If there is any damage, the manufacturer should be contacted promptly for handling.
-The installation location should be chosen indoors, avoiding direct sunlight, vibration, noise, and areas that may generate dust or pollutants.
-The workbench should be stable and can be calibrated with a spirit level. Adjust it to a horizontal state to ensure stable operation of the workbench.
2. Unpacking and Inspection
-Carefully unpack the outer packaging and inspect the workbench for any obvious transportation damage and complete accessories.
-Refer to the product manual and packing list to confirm that all accessories and parts are intact and neatly arranged.
3. Assembly process
-Assemble according to the instructions in the product manual, including installing fans, filters, workbenches, lighting fixtures, etc.
-Ensure that the installation positions of the fan and filter are correct, and the filter should be installed below the fan to ensure the circulation of clean air.
-Connect the power cord and necessary electrical components, ensuring that the voltage and current match the rated values of the ultra clean workbench.
4. Installation of filtration system
-Install pre filters, main filters (usually HEPA filters), and activated carbon filters to ensure good sealing.
-Filters should be installed in the order of airflow direction and cannot be installed in reverse.
5. Inspection and debugging
-Before connecting the power, check again whether all the assemblies are in place, whether the screws are tightened, and whether the filter is installed correctly.
-After power on, start with low voltage first to eliminate any foreign objects or obstacles that may exist inside the fan.
-Check whether the various function keys on the control panel are working properly, such as wind speed adjustment, lighting switch, etc.
6. Run tests
-Open the ultra clean workbench, adjust the wind speed to the highest, and let it run for a period of time (usually a few hours) to remove any possible pollution inside the workbench.
-Use a particle counter to detect the cleanliness of the workspace and ensure that it meets the set cleanliness level standards.
7. On site cleaning
-Before operating the workbench, the room should be cleaned to reduce the generation of dust and pollutants.
-Maintain indoor air circulation and avoid the presence of pollution sources.
8. Training and Use
-Provide necessary training to operators, including the use, maintenance, and precautions of the workbench.
-Develop usage regulations and clarify cleaning, disinfection, and maintenance procedures before and after use.
9. Regular maintenance
-Regularly inspect the performance of the workbench and verify whether the cleanliness meets the requirements.
-Replace the filter according to the manufacturer's recommended cycle to ensure air quality.
-Regularly clean and disinfect the surface of the workbench, especially the work surface and operation panel.