How to set and calibrate the temperature of the light incubator

日期:2024-03-29 09:47
浏览次数:123
摘要:I. Setting of temperature and parameters of light incubator After startup, the system enters the main monitoring screen. Press the "Stop/Run" key for 5 seconds to start or stop running, and press the "Query/Shift" key to query the set value. Self-setting function: When the compressor is normally open and in balance mode, if the temperature and humidity control is not ideal, please start the temperature and humidity self-setting. Long press the "Reduce" button to start temperature PID self-t
I. Setting of temperature and parameters of light incubator

After startup, the system enters the main monitoring screen. Press the "Stop/Run" key for 5 seconds to start or stop running, and press the "Query/Shift" key to query the set value. Self-setting function: When the compressor is normally open and in balance mode, if the temperature and humidity control is not ideal, please start the temperature and humidity self-setting. Long press the "Reduce" button to start temperature PID self-tuning.

Edit running parameters of light incubator

Press "Set" on the main interface to enter the menu, select "Running parameters" through the arrow keys, and then press "Set" to enter the editing interface. After each parameter is modified, it will automatically switch to the next parameter. If the working group is set to this group, it will run according to the working group, and if the current period is set to 0, it indicates the later program and exits editing. Fixed two ends in day and night mode. The second paragraph is set to 0 to indicate that the group is in constant mode. Long press the "Set" key to return to the main monitoring screen.